Anechoic Chambers

What Is SVSWR? Chamber Validation Above 1 GHz

How the SVSWR measurement that shows a chamber is fit for radiated emission tests above 1 GHz is made, and what affects the result.

A semi-anechoic EMC chamber is validated by NSA measurement for emission measurements below 1 GHz. But the fast digital circuits and wireless modules of today's electronic products also produce significant emissions above 1 GHz, and measuring these emissions is mandatory in many standards. At these frequencies the measurement conditions and chamber validation require a different approach.

SVSWR, the site voltage standing wave ratio measurement, is the validation method showing the chamber's suitability for emission measurements above 1 GHz. In this article we explain why a separate validation is needed, how the measurement is made, what affects the result, the time-domain method and practical advice for chamber owners.

Why a Separate Validation?

Emission measurements below 1 GHz are made in a setup where the ground reflection is included in the measurement; the antenna is moved up and down to measure the combination of the direct wave and the wave reflected from the floor. Above 1 GHz this approach changes. At these frequencies the measurement is made in an environment where the ground reflection is suppressed, that is close to free space.

For this reason, in measurements above 1 GHz, the floor between the turntable and the antenna is covered with absorbers. It must be shown that, under these conditions, reflections from the walls and floor do not affect the measurement unacceptably. Because NSA is a method that accounts for the ground reflection, it is not suitable for this purpose; the SVSWR method is used instead.

What Does Standing Wave Mean?

In a chamber, the wave leaving the transmitting antenna reaches the receiving antenna both directly and by reflecting off the walls, ceiling and floor. Depending on the position of the receiving antenna, the direct wave and the reflected waves sometimes reinforce and sometimes weaken each other. As a result, when the transmitting antenna is moved slightly, the signal level at the receiver fluctuates.

SVSWR measures the size of this fluctuation. The stronger the reflections, the larger the fluctuation. In an ideal, reflection-free environment, the signal barely changes when the transmitting antenna is moved small distances. In a real chamber some fluctuation is unavoidable; what matters is that it stays within acceptable limits.

How Is the Measurement Made?

In the SVSWR measurement, a transmitting antenna is placed at specified positions within the test volume, and the measurement antenna is kept at the position where it will be used for emission measurement. At each main position, the transmitting antenna is moved in small steps to several different points along a line towards the measurement antenna. The signal received at each point is recorded, and the difference between the highest and lowest values gives the SVSWR for that position.

The measurement is repeated at the front, left, right and centre positions of the test volume, at different heights and in both horizontal and vertical polarisation. It is made from 1 GHz up to the highest frequency the chamber will be used at, usually up to 18 GHz. Special antennas with a wide and uniform radiation pattern are used as transmitting antennas.

  • Several main positions in the test volume
  • Several measurement points along a line at each position
  • Different heights and both polarisations
  • From 1 GHz, usually up to 18 GHz

The Time-Domain Method

The classic SVSWR measurement is quite time-consuming because the transmitting antenna has to be physically moved to many positions. An alternative approach, called the time-domain method, has therefore also been added to the standards. In this method, measurements made over a wide frequency range are transformed mathematically into the time domain, and the direct signal is separated from the reflected signals.

By allowing measurements at fewer positions, the time-domain method can shorten measurement time considerably. However, it requires suitable instruments and software and experience to interpret results correctly. Which method is used depends on the laboratory's equipment and preferences; both are defined in the standard.

What Affects the Result?

The factor that most affects the SVSWR result is the chamber's absorber performance. Above 1 GHz, how well the absorbers suppress reflections shows directly in the measurement. Foam absorbers usually perform well at these frequencies; but problems can arise in chambers where only ferrite tiles are used or where the foam absorbers are very short.

The placement of floor absorbers also significantly affects the result. The absorbers must cover the area between the turntable and the antenna sufficiently, with no gaps between them. The turntable, antenna mast, cables and other objects in the chamber can also be sources of reflection. The reflection characteristics of the turntable's top surface and the antenna mast should be considered, especially at high frequencies.

Practical Advice

The frequencies and positions where SVSWR fails point to the source of the problem. High values repeated at positions facing a particular wall point to an absorber problem on that wall; high values seen at all positions over a certain frequency range may point to absorber performance in that range. This analysis shows where the fix should focus.

In chambers where floor absorbers are placed and removed for each measurement, it is very important to place the absorbers in the same arrangement every time. Marking absorber positions on the floor and documenting the placement with photographs improves measurement repeatability. Damaged absorbers should be replaced in good time.

Handover and Periodic Validation

After installation, a new semi-anechoic chamber is validated by NSA below 1 GHz and by SVSWR above 1 GHz. The results of these measurements become part of the chamber's handover documents and the laboratory's accreditation file. Validation must cover the whole test volume and frequency range in which the chamber will be used.

Because chamber performance can change over time, validation measurements are repeated at set intervals and whenever a significant change is made to the chamber. Such changes include replacing absorbers, installing a new turntable or antenna mast, or adding permanent new equipment to the chamber. Our article on NSA measurement covers NSA in detail. For chamber options, see our semi-anechoic chamber page.

Frequently Asked Questions

How long does an SVSWR measurement take?

It depends on the size of the test volume, the number of positions and the method. With the classic method it can take several days; the time-domain method can shorten this.

Is SVSWR also done above 18 GHz?

If the chamber will be used for emission measurements at higher frequencies, validation must cover these frequencies too. The scope is set by the chamber's intended use.

Must floor absorbers be permanent?

No. In many chambers floor absorbers are placed and removed for measurements above 1 GHz; but they must be placed in the validated arrangement every time.

What is the SVSWR limit?

According to CISPR 16-1-4, the SVSWR measured between 1 and 18 GHz must not exceed 6 dB.

What is the difference between NSA and SVSWR?

NSA is the validation that accounts for the ground reflection and is used below 1 GHz. SVSWR assesses wall reflections for measurements above 1 GHz where the ground reflection is suppressed.

What is done if SVSWR fails?

The frequencies and positions with high values are analysed, absorber, floor absorber placement or reflective object issues are fixed and the measurement is repeated.

Let's Find the Right Solution for Your Project

Your free site survey and quote will be ready within 24 hours.

Get a Quote

Get a Quote

Step 1 / 2 · Your Project

What kind of room do you need?
Approximate size