What Is NSA? Validating an Anechoic Chamber to CISPR 16
What the NSA (normalized site attenuation) measurement is and how it is done to show a semi-anechoic chamber is fit for emission testing.
We compare the semi-anechoic chamber, with its reflective floor, and the fully anechoic chamber, with absorbers on every surface, and where each is used.
When people think of an anechoic chamber, they usually picture a room with walls covered in pyramid-shaped absorbers. But there are two basic types of these chambers, and the difference between them goes far beyond whether the floor has absorbers. This difference directly determines which tests can be done in the chamber, which standards can be met and how much the chamber costs.
In this article we explain the difference between a semi-anechoic chamber (SAC) and a fully anechoic room (FAR), which measurements each is designed for, which chamber to choose for which test, whether both needs can be met in a single chamber and how this choice affects budget and building.
In a semi-anechoic chamber the walls and ceiling are lined with absorbers, but the floor is conductive and reflective. This structure imitates the ideal open area test site on which EMC emission limits are based. On an open area site only the ground reflects; a semi-anechoic chamber likewise suppresses wall and ceiling reflections while keeping the ground reflection.
Most EMC emission measurements are made in this type of chamber according to standards such as CISPR 32 and CISPR 11. Depending on the test distance, there are 3 m, 5 m and 10 m chambers. During measurement the device is rotated on a turntable, and the receiving antenna is moved up and down between 1 m and 4 m to find the maximum signal combined with the ground reflection. The chamber's suitability is validated by NSA measurement. For details see our semi-anechoic chamber page.
In a fully anechoic room, all surfaces including the floor are lined with absorbers. This structure imitates free space with no reflections at all. Applications such as antenna measurements, radar cross-section measurements and over-the-air performance tests of wireless devices are carried out in this type of room because they need a reflection-free environment.
In fully anechoic rooms, absorbers are usually pyramid-shaped and sized according to the lowest frequency the room will operate at. Because the floor is also lined with absorbers, removable walkways or platforms are placed on top of the absorbers so people can walk and carry equipment. Our fully anechoic room options are on our fully anechoic room page.
The choice is made according to the tests to be performed and the standards to be met. For EMC emission tests aimed at product certification, a semi-anechoic chamber is usually the only option. Radiated immunity tests are also mostly carried out in a semi-anechoic chamber, with additional absorbers placed on the floor in front of the test area. For emission measurements above 1 GHz, the floor is also temporarily covered with absorbers.
Organisations doing antenna design, wireless device performance and radar measurements need a fully anechoic room. In these measurements, a ground reflection is unacceptable because it would distort the antenna's radiation pattern. We explain the measurement methods in detail in our articles on OTA testing: TRP and TIS and the antenna measurement chamber.
The absorber requirements of the two chamber types differ. Because EMC chambers make measurements starting from low frequencies such as 30 MHz, ferrite tiles or hybrid absorbers that stay compact at low frequencies are preferred. In antenna measurement rooms, the frequencies used are usually higher and long pyramid absorbers perform better.
Absorber selection directly affects chamber size. For the same test distance, a fully anechoic room lined with long pyramids usually requires larger outer dimensions than a ferrite-lined semi-anechoic chamber. We explain in detail the differences between absorber types and the frequencies at which they are effective in our article on ferrite tiles and pyramid absorbers.
For organisations wanting to do both EMC and antenna measurements, using a single chamber instead of building two is an attractive option. The floor of a semi-anechoic chamber can be covered with removable absorbers to make it temporarily fully anechoic. This approach offers an adequate solution for some measurements.
But this flexibility brings some limitations. Placing and removing floor absorbers for each type of measurement takes time, the absorbers can be damaged in the process and the chamber's performance in each mode may not be as good as that of a chamber designed only for that purpose. For organisations with high measurement volume and frequency, two separate chambers are usually more efficient.
Which is more expensive depends on chamber size and absorber type. Comparing a semi-anechoic chamber with a fully anechoic room of the same size, covering the floor with absorbers in the fully anechoic room adds extra cost. But semi-anechoic EMC chambers also require a significant investment because of the turntable, antenna mast and low-frequency absorbers.
In terms of the building, both chamber types need considerable space and height. A 10 m semi-anechoic chamber can be too large to fit in many buildings and may require a separate structure. Ferrite tiles also add significant weight. It is therefore recommended to finalise the chamber type only after assessing what the building allows. We have collected building preparation items in our article on building preparation for a shielded room.
The validation of the two chamber types also differs. Semi-anechoic chambers are validated by NSA measurement below 1 GHz and by SVSWR measurement above 1 GHz. These measurements prove that the chamber meets the conditions defined in EMC standards. For radiated immunity tests, a field uniformity measurement is also made.
In fully anechoic rooms, quiet zone performance is measured. This measurement shows how low a reflection level the room provides in the region where the antenna under test is located. In both cases the validation results become part of the chamber's handover documents and the laboratory's quality system. For details see our article on NSA measurement.
Some antenna measurements can be made by temporarily covering the floor with absorbers. However, a fully anechoic room is more suitable for measurements requiring high accuracy.
Some standards allow it under certain conditions, but most EMC emission tests require a semi-anechoic chamber. Which standard applies should be checked in advance.
It depends on chamber size and absorber type. Floor absorbers add cost in a fully anechoic room, while semi-anechoic EMC chambers need a turntable and low-frequency absorbers.
SAC stands for semi-anechoic chamber, and FAR for fully anechoic room.
It depends on product size and the standard to be applied. For small products a 3 m chamber is usually enough; large products and some standards may require a 10 m chamber.
It can be economical for organisations with low measurement volume. For organisations measuring frequently, two separate chambers are usually more efficient.