Anechoic Chambers

What Is NSA? Validating an Anechoic Chamber to CISPR 16

What the NSA (normalized site attenuation) measurement is and how it is done to show a semi-anechoic chamber is fit for emission testing.

For an emission measurement made in a semi-anechoic chamber to be reliable, you need to be sure that the chamber itself does not distort the result. Inadequate absorbers, wrongly chosen chamber dimensions or a problem with the continuity of the ground plane lead to the same device giving different results in different chambers.

The normalised site attenuation (NSA) measurement is the basic validation method showing how closely an EMC chamber behaves like an ideal open area test site. In this article we explain why chambers are validated, how the NSA measurement is made, the SVSWR method used above 1 GHz, how measurement results are evaluated and how often validation should be repeated.

Why Is the Chamber Validated?

EMC emission limits are set assuming an ideal open area test site (OATS). This site has only a conductive ground plane and no other objects nearby to cause reflections. In real life it is very hard to find such a site and keep it free of outside radio signals, so measurements are made in semi-anechoic chambers that imitate an open area.

How well the chamber imitates an open area depends on absorber performance, chamber dimensions and geometry. The validation measurement proves that the chamber does this within an acceptable tolerance. Results of measurements made in an unvalidated chamber are not accepted by accredited laboratories and regulators. Chamber validation is defined in the CISPR 16-1-4 standard.

How Is NSA Measured?

Two antennas are used in the NSA measurement: one transmitting and one receiving. The transmitting antenna is placed where the device under test will be in the chamber, and the receiving antenna at the measurement distance, 3 m or 10 m away. The receiving antenna is moved up and down over a specified height range to capture the effect of the ground reflection, and the maximum signal is recorded at each frequency.

The measured signal loss, after the antennas' own characteristics are removed, is compared with the value calculated theoretically for an ideal open area site. The difference shows how far the chamber deviates from the ideal site. The measurement is made in both horizontal and vertical polarisation at frequencies between 30 MHz and 1 GHz.

Test Volume and Measurement Positions

During a test, a device is rotated on the turntable, so different parts of the device radiate from different positions. The chamber must perform well not only at the centre of the turntable but throughout the volume the device can occupy. This volume is called the test volume.

For this reason the NSA measurement is not made at a single point but at several positions representing the test volume. The transmitting antenna is placed at the centre of the test volume, at its front, rear, left and right edges and at different heights, and the measurement is repeated. The larger the test volume, the more measurements are needed, and the chamber must meet the tolerance throughout this volume.

  • Centre of the test volume
  • Front, rear, left and right edge positions
  • Different transmitting antenna heights
  • Horizontal and vertical polarisation

Above 1 GHz: SVSWR

At frequencies above 1 GHz, emission measurements are made with the ground reflection suppressed; for this, absorbers are placed on the floor between the turntable and the antenna. To validate the chamber at these frequencies, the site voltage standing wave ratio (SVSWR) method is used instead of NSA. This method measures how much reflections from the chamber walls cause the signal in the test volume to fluctuate.

In the SVSWR measurement, a transmitting antenna is moved in small steps to different positions in the test volume, and the received signal is recorded at each position. The stronger the reflections, the larger the fluctuation in the signal. CISPR 16-1-4 requires this fluctuation to stay below a certain limit. We explain the measurement in detail in our article what is SVSWR measurement.

What If NSA Fails?

Exceeding the tolerance in the NSA measurement means the chamber does not perform adequately at some frequencies or positions. The source of the problem is usually frequency regions where absorbers fall short, gaps between ferrite tiles, surfaces not covered with absorbers or reflective objects in the chamber. The turntable, antenna mast and cable routes can also contribute to reflections.

Knowing at which frequency and at which position the problem appears helps find the source. For example, a deviation repeated at positions facing a particular wall points to an absorber problem on that wall. After the correction, the relevant measurements are repeated to verify that the chamber meets the tolerance. We explain how absorber performance affects the chamber in our article on ferrite tiles and pyramid absorbers.

Is NSA Needed for a Pre-Compliance Chamber?

Pre-compliance chambers are used not for certification measurements but for finding problems at the design stage. NSA validation is therefore not mandatory. However, knowing the chamber's performance contributes greatly to interpreting pre-test results. A simple NSA measurement shows at which frequencies the chamber gives reliable results and at which frequencies care is needed.

For pre-compliance chambers that may be used for accredited measurements in future, it is recommended to design the chamber from the start to meet the NSA tolerance. Meeting this tolerance later through absorber and geometry changes is usually more costly. For our pre-compliance chamber options, see our SAC-3 page.

Handover and Periodic Validation

A new EMC chamber is validated with NSA and, where needed, SVSWR measurements after installation, and the measurement report becomes part of the chamber's handover documents. This report is used to prove the chamber's suitability in the laboratory's accreditation process and in customer audits.

The chamber's performance can change over time. Damaged absorbers, new equipment added to the chamber or changes to the turntable and antenna mast can affect validation results. Accredited laboratories therefore repeat validation measurements at intervals defined in their quality system. For our large-scale test chamber options, see our SAC-10 page.

Frequently Asked Questions

How long does an NSA measurement take?

It depends on the chamber size, test volume and number of measurement positions. It is usually completed in one to a few days.

Is NSA needed for a pre-compliance chamber?

It is not mandatory, but knowing at which frequencies the chamber gives reliable results helps interpret pre-test results.

What is done if NSA fails?

The frequencies and positions where deviations appear are examined, absorber or reflective object issues are fixed and the relevant measurements are repeated.

What is the NSA tolerance?

According to CISPR 16-1-4, the measured NSA must deviate from the theoretical value by at most ±4 dB.

Why is a different method used above 1 GHz?

At these frequencies measurements are made with the ground reflection suppressed. The SVSWR method, which measures wall reflections, is used to validate the chamber.

How often should validation be repeated?

At intervals defined in the laboratory's quality system and whenever a significant change is made to the chamber.