IEC 60601-1-2: EMC Requirements and Risk for Medical Devices

The test levels of IEC 60601-1-2, the EMC standard for medical electrical equipment, which vary by environment, and how it ties into risk management.

In a hospital room, patient monitors, infusion pumps, ventilators and imaging equipment work in the same environment as nurses' radios and patients' mobile phones. Medical devices used at home have to work next to microwave ovens, wireless networks and household appliances. In these environments, a medical device malfunctioning because of electromagnetic interference can directly threaten patient safety.

IEC 60601-1-2 is the basic standard for the electromagnetic compatibility of medical electrical equipment. In this article we explain why EMC is treated differently for medical devices, how levels are set according to the environment of use, proximity tests for wireless devices, the link with risk management, the importance of the test plan and the test infrastructure.

Why Is EMC Different for Medical Devices?

In an ordinary electronic product, electromagnetic interference usually causes a temporary malfunction that annoys the user. In medical devices, the same interference could cause a patient to receive the wrong dose of a drug, an alarm not to sound or a life-support device to stop. That is why EMC in medical devices is treated as a direct part of patient safety.

IEC 60601-1-2 builds this approach on two basic concepts: basic safety and essential performance. Basic safety means the device does not create an unacceptable risk to the patient or user even under electromagnetic interference. Essential performance means the device maintains those functions whose loss or degradation would create an unacceptable risk. For example, a patient monitor's alarm function may be part of essential performance.

Levels by Environment of Use

The current edition of the standard sets test levels according to the environment in which the device will be used. Different environments of use are defined: the professional healthcare facility environment, meaning hospitals and clinics; the home healthcare environment, meaning homes, vehicles and outdoor areas; and special environments. Because each environment has different electromagnetic characteristics, immunity test levels vary accordingly.

Levels for the home healthcare environment are generally higher. This is because the electromagnetic environment in homes is not controlled and devices do not operate in an environment managed by technical staff, as in hospitals. For example, in the radiated RF immunity test a higher field strength is applied for the home environment than for the professional healthcare facility environment.

  • Professional healthcare facility environment: hospitals and clinics
  • Home healthcare environment: homes, vehicles, outdoor areas
  • Special environments: such as military, heavy industry or medical treatment areas

Proximity to Wireless Devices

Medical devices can be used very close to transmitters such as mobile phones, two-way radios and wireless network equipment. The standard defines special proximity tests to represent this situation. In these tests the device is exposed to high-intensity fields generated from a short distance at common wireless communication frequencies. These fields are much higher than those in the general radiated immunity test.

With the latest update of the standard, proximity tests to low-frequency magnetic field sources have also been added. These tests represent sources such as RFID readers, wireless chargers and anti-theft systems. Because these sources are becoming more common in hospitals and everyday life, it is important that medical devices are immune to them.

The Link with Risk Management

IEC 60601-1-2 links the EMC assessment to the device's risk management process. The manufacturer must assess the environment where the device will be used, possible electromagnetic interference in that environment and the effects of this interference on the device's basic safety and essential performance. This assessment is made in line with ISO 14971, the risk management standard for medical devices.

The risk assessment may require test levels higher than those recommended by the standard. For example, if there are powerful transmitters in the environment where the device will be used, the standard's minimum levels may not be enough. Whether a degradation seen in the test is acceptable is also decided according to the risk assessment.

Test Plan and Documentation

IEC 60601-1-2 explicitly requires a detailed test plan to be prepared before testing. This plan contains the device description, environment of use, essential performance, test levels, operating modes during the test, the functions to be monitored and pass criteria. The plan must be consistent with the risk management file. We cover general test plan preparation in our article on the EMC test plan.

The standard also requires EMC-related information to be given in the device's user documentation. This includes the environment the device is designed for, precautions the user should take and warnings about using portable RF communication equipment near the device. These documents help the user operate the device safely.

Emission Requirements

The emissions of medical devices are assessed according to CISPR 11. Devices are divided into groups and classes according to their environment of use. Devices used in the home healthcare environment usually have to meet Class B limits. Some devices used in the professional healthcare facility environment can be assessed against Class A limits.

In medical devices, emission is not just a regulatory requirement but also important for the safety of other medical devices. Noise emitted by one device can disturb the operation of a sensitive monitor or measuring instrument nearby. We discuss interference sources in hospitals in our article on EMI sources in hospitals.

Test Infrastructure

IEC 60601-1-2 testing adds some special requirements to general EMC test infrastructure. Proximity tests require signal sources, amplifiers and antennas able to generate high field strengths at specific frequencies. These tests are usually carried out in a semi-anechoic chamber or a shielded room. Special coils are used for magnetic field proximity tests.

For medical device manufacturers, an in-house pre-compliance room is a big advantage for finding problems early in design. Proximity tests in particular can be done early in the design and weak points fixed before the certification test. We explain medical device test strategy in our article on medical device EMC test strategy and our healthcare solutions in our healthcare and medical solutions.

Frequently Asked Questions

Why are home-use medical devices tested more strictly?

The electromagnetic environment in homes is not controlled and devices do not operate in an environment managed by technical staff. Immunity levels for the home healthcare environment are therefore generally higher.

What is essential performance?

Those functions of the device whose loss or degradation would create an unacceptable risk. The manufacturer determines these functions through the risk management process.

Do proximity tests need a special chamber?

A semi-anechoic chamber or shielded room is usually used. Because high field strengths are generated, the test must be done in a controlled environment.

Which standard are medical device emissions measured to?

Medical device emissions are assessed according to CISPR 11, with the group and class suited to the environment of use.

Can test levels be set higher than the standard?

Yes. The risk assessment may require higher levels because of the environment where the device will be used.

Is EMC information needed in user documentation?

Yes. The standard requires information such as the environment of use, precautions to take and warnings about portable RF devices to appear in the documentation.