Magnetic resonance (MR) imaging creates images by measuring extremely weak radio-frequency signals coming from inside the body. These signals are so weak that noise from a radio transmitter outside the room, a two-way radio or even an electronic device can appear in the image as lines and distortions.
The RF cage in an MR room is the shielded structure that prevents these outside signals from reaching the imaging area. In this article we explain why MR scanners are shielded, the difference between the RF cage and magnetic shielding, how doors, windows and penetrations are designed, the earthing and insulation requirements of the cage and why acceptance and periodic measurements are critical.
An MR scanner places the patient in a strong magnetic field and excites hydrogen atoms in the body with radio waves at a specific frequency. When the excitation ends, the atoms emit very weak signals at the same frequency, and the scanner receives these signals to build the image. This frequency is proportional to the magnet strength: about 64 MHz for a 1.5 tesla scanner and about 128 MHz for a 3 tesla scanner.
These frequencies are close to the bands used by FM radio, two-way radio and other communication systems. When an outside signal reaches the MR receiver, it creates noticeable lines or distortions in the image. This lowers diagnostic quality and may require scans to be repeated. The RF cage preserves image quality by keeping these outside signals out of the room.
Two different shielding concepts are often confused in MR rooms. The RF cage protects against radio-frequency signals and is a continuous conductive structure, usually made of copper or galvanised steel panels covering all surfaces of the room. Its purpose is to keep outside signals from getting in and the scanner's own RF signals from getting out.
Magnetic shielding, on the other hand, is used to limit how far the strong static magnetic field created by the magnet extends outside the room. This field can pose a risk to people with pacemakers and to sensitive equipment nearby. Magnetic shielding is usually made of thick steel plates, and whether it is needed is calculated by physicists according to the scanner type and the room's surroundings. An MR room may need both, but their designs and purposes are completely different.
The RF cage is built to form a continuous conductive surface across the floor, walls and ceiling of the room. Panels are bonded to each other electrically and continuously, and joints are designed so as not to create leaks. The cage floor is insulated from the building concrete, so that the cage is earthed at only one point, in a controlled way.
Insulating the cage from the building earth is an important requirement specific to MR rooms. If the cage touches the building at more than one point, ground loops can form, and these loops can cause noise in the image. For this reason the insulation resistance of the cage is measured during and after installation, and the single-point earth connection is documented.
The MR room door is the most critical and most used part of the cage. It must be wide enough for hospital beds and stretchers to pass easily, while fully compressing its gaskets at every closing to keep the shielding continuous. Door gaskets need regular maintenance and replacement because of heavy use.
The observation window between the control room and the MR room is made of special glass with a fine metal mesh inside. This glass stops RF signals without impairing the view. Medical gases, ventilation and equipment cables enter the room through waveguide penetrations, honeycomb filters and RF filters. The emergency quench pipe of the magnet's cooling system is also routed out through the cage via a special penetration.
After the RF cage is installed, its shielding performance is measured before the MR scanner is commissioned. The measurement is made over the frequency range specified by the scanner manufacturer and at critical points such as the door, window and filter panel. The results are used to document that the manufacturer's installation requirements are met and are added to the hospital's technical documentation.
The cage's performance can decrease over time. Door gaskets wear, later refurbishments open new penetrations and connections loosen. When unexplained lines or noise appear in images, the RF cage is one of the first places to check. The condition of the cage should therefore be checked during periodic shielding measurement and the results recorded.
When an MR scanner is replaced, it should be assessed whether the existing RF cage is sufficient for the new scanner. Especially when a 3 tesla scanner replaces a 1.5 tesla one, the operating frequency rises and it must be measured whether the cage performs adequately at this frequency. In some cases the cage can be made compatible with the new scanner by renewing the door and filters.
In cage refurbishment projects, the hospital's continuity of service is very important. Work is therefore planned in line with the scanner replacement schedule and completed as quickly as possible. Door gasket replacement and small repairs can usually be done in a single day. For our maintenance and repair services, see our maintenance and measurement page.
MR rooms are the best-known example of shielding needs in hospitals, but not the only one. EEG and sleep laboratories, electrophysiology rooms and some research laboratories also need protection from outside noise. Because each has a different frequency range and sensitivity, the shielding design varies with the need.
We explain the main interference sources in the hospital environment in our article on EMI sources in hospitals, and requirements specific to EEG rooms in our article on EEG room shielding. Our solutions for the healthcare sector are on our healthcare and medical page.
No. An MR room needs high RF shielding around the scanner's operating frequency; in an EEG room, low-frequency mains noise is more decisive. The designs differ accordingly.
Yes. Performance can be recovered by renewing the door, gaskets and filters or repairing the cage. The scope is determined by measurement.
It depends on the size of the room and the number of measurement points; most MR rooms are measured within a day.
Because the operating frequency rises, the cage needs to be measured at this frequency. Depending on the result, the door and filters may need renewing.
No. The RF cage protects against radio-frequency signals. Separate magnetic shielding is designed to limit the magnet's static field.
It is one of the possible causes. Door gaskets, filters and penetrations opened later should be checked and the cage measured; scanner-related causes should also be assessed separately.