Why Partial Discharge Testing Needs a Shielded Lab
Why partial discharge measurement in cable, transformer and insulator testing is so sensitive to outside noise, and the role of a shielded lab.
How the RF filters on every power line entering a shielded room are chosen, leakage current and safety issues, and common mistakes.
However good the walls of a shielded room are, every power cable entering it is an antenna that can carry outside signals in. Every line feeding the room's lighting, sockets, air conditioning and test equipment is therefore passed through a power line filter at the point where it enters the room. Filter selection directly affects the safety of the people working inside as much as the room's performance.
In this article we explain how power line filters work, how current, voltage and attenuation ratings are chosen, why leakage current matters, how filters should be mounted and the most common mistakes.
A power cable outside a shielded room collects radio signals from the environment and carries them along the cable. If the cable enters the room unfiltered, these signals are re-radiated inside the room and largely cancel the benefit of the shielding. Likewise, signals generated inside the room can leave via the cable; in security rooms this means a risk of information leakage.
A power line filter lets energy at mains frequency through while stopping high-frequency signals. These filters are usually multi-stage circuits made of inductors and capacitors. The filter is mounted on the wall of the shielded room and its body is in continuous electrical contact with the shield, so the signals it stops are diverted to earth via the outer surface of the shield.
The first criterion in filter selection is the current the filter will carry and the voltage it will operate at. The filter's rated current should be chosen to cover the total current of all loads in the room, inrush currents and future load growth. A filter with an insufficient current rating heats up, has a shorter life and can create a safety risk.
The voltage rating must match the supply feeding the room. Different filters are used for single-phase, three-phase and direct current supplies. In some test chambers, separate filters are planned for equipment needing supplies at different voltages and frequencies. Loads such as UPS outputs and frequency converters also need special attention in filter selection.
A filter's performance is expressed as how much it attenuates signals over a given frequency range. Shielded room filters are usually designed to provide high attenuation over a wide range, from a few kilohertz up to the gigahertz range. The filter's attenuation must match the room's target shielding value; otherwise the filter becomes the room's weakest link.
The attenuation values given by filter manufacturers are usually obtained in a standard measurement setup. In real use, performance can differ depending on how the filter is mounted and on source and load impedances. The performance of filters should therefore also be verified by measurements in the filter area during the room's acceptance measurement.
The capacitors in power line filters are connected between the phase and neutral lines and earth. This causes a continuous leakage current to flow into the earth line even at mains frequency. In rooms using several filters, these leakage currents add up and can cause standard residual current devices to trip unintentionally.
Filter selection should therefore be made together with the room's electrical protection arrangement. The filters' leakage current values, the characteristics of the protection devices on the room's supply and the relevant electrical regulations should be evaluated together. Continuity of the earth connection is vital: a dangerous voltage can appear on the body of a filter whose earth connection has broken.
A filter's performance depends largely on how it is mounted. The filter must be mounted on the shielded room's wall so that its body makes continuous, low-impedance contact with the shield. Paint or coating on the mounting surface should be cleaned off and a conductive gasket used if necessary. The filter's input side, the dirty side, must be outside the room and its output side, the clean side, inside.
It is also important not to bring dirty-side and clean-side cables close to each other. If unfiltered cables outside the room run in the same cable tray as, or next to, filtered cables inside the room, signals can bypass the filter and jump from one cable to the other. Filter inputs are therefore usually gathered in a separate filter box outside the room.
The most common mistake is choosing the filter's current rating only for today's load. An air conditioner or test instrument added to the room later can exceed the filter's capacity, and adding a new filter requires opening up and re-verifying the room. Filter capacity should therefore be chosen with a margin for future load growth.
The second common mistake is not taking leakage current into account in the electrical design. If residual current devices keep tripping after installation, users may disable the protection devices, which is a serious safety risk. The third mistake is trying to run data and control cables through power filters; separate signal filters or fibre optic connections should be used for these cables. See our article on building preparation.
Power line filters work trouble-free for many years when chosen and mounted correctly. But during periodic maintenance, the tightness of filter connections, continuity of the body's contact with the shield, the earth connection and any signs of overheating of the filter body should be checked. Filters' insulation resistance and leakage current values can also be measured at intervals.
When a filter needs replacing, the new filter's current, voltage and attenuation ratings must match or exceed those of the old one. After replacement, a check measurement in the filter area should verify that the room's performance is maintained. For filter replacement and room modifications, see our maintenance and modification page, and for our shielded room solutions, our RF/EMI shielded room page.
No. A filter is needed for every power line entering the room; once a line is filtered, it can be distributed to several sockets inside the room.
When chosen and mounted correctly they are long-lasting. Overload, poor cooling or voltage surges can shorten their life; periodic checks are recommended.
Yes, but not power filters; special filters suited to the signal line are used. Fibre optic connections are preferred wherever possible.
Filter capacitors continuously leak a small current to earth. When the leakage currents of several filters add up, they can trip standard devices; the protection arrangement should be designed accordingly.
The total current of all loads in the room, inrush currents and a margin for future load growth are evaluated together.
Yes. A check measurement in the filter area verifies that the room's shielding performance is maintained.